CVE-2023-25931

Medium 6.8No known exploitationFix available

Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue

Record last updated

CVSS score

6.8

Medium on a 0 to 10 scale

EPSS, chance of exploitation in 30 days

0.3%

Higher than 19% of scored CVEs. A prediction, not evidence.

CISA remediation deadline

None

Not in the KEV catalog

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Weakness: CWE-620

Patching and mitigation

Patch status
Fix available (source: CVE record (Medtronic))
Fix documented (source date)
Vendor solution
Current versions of the application has mitigated this vulnerability. Please refer to the Medtronic Security Bulletin for update guidance.
Mitigation
None documented
Vendor guidance
medtronic.com (opens in a new tab)

Affected products

VendorProductAffected versionsFixed inPer
MedtronicInterStim Applicationsfrom InterStim X Clinician before A51300A51300CVE Program
MedtronicInterStim Applicationsfrom Micro Clinician before A51200A51200CVE Program
medtronicinterstim_x_cliniciana51300Not documentedNIST NVD
medtronicmicro_cliniciana51200Not documentedNIST NVD

Timeline

  1. Disclosure

    CVE record published by Medtronic

    Source: CVE Program (opens in a new tab)

  2. Patch

    Fix documented by CVE record (Medtronic)

    Source: CVE record (Medtronic) (opens in a new tab)