CVE-2023-25931
Medium 6.8No known exploitationFix availableMedtronic Micro Clinician & InterStim X Clinician App Password Reset Issue
Record last updated
CVSS score
Medium on a 0 to 10 scale
EPSS, chance of exploitation in 30 days
Higher than 19% of scored CVEs. A prediction, not evidence.
CISA remediation deadline
Not in the KEV catalog
Description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Weakness: CWE-620
Patching and mitigation
- Patch status
- Fix available (source: CVE record (Medtronic))
- Fix documented (source date)
- Vendor solution
- Current versions of the application has mitigated this vulnerability. Please refer to the Medtronic Security Bulletin for update guidance.
- Mitigation
- None documented
- Vendor guidance
- medtronic.com (opens in a new tab)
Affected products
| Vendor | Product | Affected versions | Fixed in | Per |
|---|---|---|---|---|
| Medtronic | InterStim Applications | from InterStim X Clinician before A51300 | A51300 | CVE Program |
| Medtronic | InterStim Applications | from Micro Clinician before A51200 | A51200 | CVE Program |
| medtronic | interstim_x_clinician | a51300 | Not documented | NIST NVD |
| medtronic | micro_clinician | a51200 | Not documented | NIST NVD |
Timeline
Disclosure
CVE record published by Medtronic
Source: CVE Program (opens in a new tab)
Patch
Fix documented by CVE record (Medtronic)
Vendor guidance and references
- https://www.medtronic.com/en-us/e/product-security/security-bulletins/pelvic-health-interstim-micro.html (opens in a new tab)vendor-advisory
- https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html (opens in a new tab)vendor advisory, x_transferred
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-061-01 (opens in a new tab)third-party-advisory